ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)

FDA Recall Z-0902-2015 · Posted 2015-01-02

Recall StatusTerminated
Product CodeJJE
Date Initiated2014-11-13
Date Terminated2015-03-19
Product Quantity443 units

Reason for Recall

HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes

Firm Action

A recall letter was sent on 11/13/14 to all customers who purchased the ABX PENTRA 400 Clinical Chemistry Analyzer. The letter informs the customer of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to contact their local HORIBA Medical representative.

Root Cause

Package design/selection

Distribution Pattern

Nationwide in US