Medtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090, Size: 5MM X 40MM X 90CM; 4) AB14W050040150, Size: 5MM X 40MM X 150CM; 5) AB14W050060150, Size: 5MM X 60MM X 150CM; 6) AB14W050080090, Size: 5MM X 80MM X 90CM; 7) AB14W050080150, Size: 5MM X 80MM X 150CM; 8) AB14W050100150, Size: 5MM X 100MM X 150CM; 9) AB14W050120150, Size: 5MM X 120MM X 150CM; 10) AB14W050150150, Size: 5MM X 150MM X 150CM; 11) AB14W050200090, Size: 5MM X 200MM X 90CM; 12) AB14W050200150, Size: 5MM X 200MM X 150CM; 13) AB14W060040090, Size: 6MM X 40MM X 90CM; 14) AB14W060040150, Size: 6MM X 40MM X 150CM; 15) AB14W060060150, Size: 6MM X 60MM X 150CM; 16) AB14W060080150, Size: 6MM X 80MM X 150CM; 17) AB14W060100150, Size: 6MM X 100MM X 150CM; 18) AB14W060120150, Size: 6MM X 120MM X 150CM; 19) AB14W060150150, Size: 6MM X 150MM X 150CM; 20) AB14W060200150, Size: 6MM X 200MM X 150CM; Balloon Dilatation Catheter

FDA Recall Z-0904-2025 · Posted 2025-01-15

Recalling Firmev3 Inc
Recall StatusOpen, Classified
Product CodeLIT
Date Initiated2024-12-10
Product Quantity5189 units

Reason for Recall

There is the potential for in-vivo balloon deflation issues, including difficulty or inability to deflate the balloon in specific lots.

Firm Action

Medtronic issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 12/10/2024 via mail courier. The notice explained the issue, potential harms, and requested the following actions be taken: "" Immediately locate and quarantine all affected and unused NanoCross devices as listed in the provided scope. " Return all unused affected products in your inventory to Medtronic. Contact Medtronic Vascular Customer Service at 800-716-6700 to initiate a product return. Your local Medtronic Representative can assist you with the initiation of the return. " Complete and return the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information, and return it to [email protected] within 30-days of receipt of this letter. " Please forward this notice to all those who need to be aware within your organization and maintain a copy for your records."

Root Cause

Process change control

Distribution Pattern

Worldwide distribution - US (& PR) Nationwide and the countries of Chile, Cyprus, Estonia, Germany, Italy, Japan, Jordan, Puerto Rico, Spain, United States, Uzbekistan, and Yemen.