Coulter LH750 Hematology Analyzer. Part Number 6605632.
FDA Recall Z-0905-04 · Posted 2004-07-20
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeGKZ
Date Initiated2004-03-31
Date Terminated2012-04-30
Product Quantity981
Reason for Recall
If a customer enters a dilution factor for Sample B prior to the results from Sample A being transfered to the workstation, the predilute multiplication factor will be applied erroneously to the Sample A's results once they are received at the workstation.
Firm Action
A Product Corrective Action Letter was mailed to all customers who currently have the instrument in thier lab. The letter includes a description of what conditions need to be met for the error to occur and the proper steps to prevent the issue from occuring. The letter was sent by US Mail on March 31, 2004.
Root Cause
Other
Distribution Pattern
Nationwide and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement