Separation Technology Creamatocrit Plus, Catalog Number: 100-146
FDA Recall Z-0905-2022 · Posted None
Advertisement
Recalling FirmStanbio Laboratory, LP
Recall StatusTerminated
Product CodeQEI
Date Initiated2022-03-16
Date Terminated2025-01-02
Product Quantity50
Reason for Recall
The device was being marketed as an IVD without a 510(k). Therefore, performance characteristics of these breast milk analyzers have not been adequately validated. An erroneous result could lead to sub-optimal nutrition supplementation for patients (infants)
Firm Action
On 03/16/22, recall notices were mailed to customers who were asked to take the following actions:
Stop using the affected device while you review the new Research Use Only labeling included with this letter. If you are using the affected device for research use only, then apply/use the new labeling. If you are using the affected human breast milk analyzer to provide information relating to the nutritional management of infants, then destroy the product in accordance with local regulations.
If the affected product was further distributed or forwarded elsewhere, then forward this recall notice to all that need to be aware; alternately, provide a customer contact list so the recalling firm can send this recall notice to these customers.
Please complete the Recall Return Response form to the recalling firm. Customers with additional questions are asked to contact Technical Support, Monday through Friday, 8:00 AM to 5:00 PM, Central Time at 1-800-531-5535 or via email at [email protected].
Root Cause
No Marketing Application
Distribution Pattern
US Nationwide Distribution CA, TX, OK, SD, IL, NV, MO, FL, WA, KY, WI, PA, NC, GA, MD, MN
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.