c.cam Gamma Camera, Model 9VIR1200; Siemens Medical Solutions USA, Inc., Molecular Imaging, 2501 N. Barrington Road, Hoffman Estates, IL 60195-5203; Made in Denmark;

FDA Recall Z-0906-2007 · Posted 2007-06-06

Recall StatusTerminated
Product CodeKPS
Date Initiated2007-04-30
Date Terminated2009-01-09
Product Quantity413 cameras

Reason for Recall

Chair Breakage: Under extreme load, the chair back of the c.cam emission computed tomography system may bend and eventually break. This could be caused by the accumulated stress from heavy patients seating or stretching themselves in the chair or simply from long time use.

Firm Action

Important Safety Notice letters dated 4/30/07 on Siemens letterhead were sent to the end users, informing them of the potential risk that the c.cam chair may bend or break under heavy load. They were told that a Siemens service representative will contact them within the next few days to arrange the repair of the chair at no cost to the end user. To ensure the chair does not become damaged, SIemens recommended that the c.cam not be used for patient scanning until the update has been performed on the chair. Any questions regarding the Important Safety Notice were directed to their service representative or Siemens' office at: 1-800-888-7436 (USA).

Root Cause

Other

Distribution Pattern

Worldwide-USA, Canada, France, Germany, Israel, The Netherlands and Portugal.