Terminated

SIEMENS brand ARTISTE syngo RT Therapist (RTT) version 4.3, ARTISTE Model, Part No. 08162815, Digital linear accelerators, MOSAIQ OIS system, Distributed by SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Manufactured by SIEMENS AG, MEDICAL SOLUTIONS, GERMANY.

FDA Recall Z-0907-2013 · Posted 2013-03-05

Recall StatusTerminated
Product CodeIYE
Date Initiated2013-01-17
Date Terminated2013-03-05
Product Quantity5 units

Reason for Recall

Siemens became aware of a potential safety issue when using the "In-session resumption" feature of the syngo RT Therapist (RTT) 4.3 software. The RTT will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.

Firm Action

The firm issued "Urgent Medical Device Correction" letters via mail to all its direct consignees, informing them of the affected product issue and providing instructions on the recall. The firm stated that the release of this UI TH005/13/S to distribute the Customer Safety Advisory Notice began on January 17, 2013.

Root Cause

Software design

Distribution Pattern

5 devices, all located outside of the United States in Germany.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.