Terminated

Low-Boy Bed, AC-powered adjustable hospital bed with a low position of 7'' to a high position of 30''. The beds are distributed by SIZEWise Rentals, Ellis, KS.

FDA Recall Z-0908-06 · Posted 2006-05-25

Recalling FirmRayes Inc
Recall StatusTerminated
Product CodeFNL
Date Initiated2005-11-14
Date Terminated2007-12-20
Product Quantity161 beds

Reason for Recall

The bed may drop due to a malfunction of the actuator exceeding its stop.

Firm Action

The recalling firm met with their one customer on/about 11/14/05 explaining the reason for correction. The customer then notified their company stores and dealers via service bulletin dated 2/28/06 that they would be receiving new acutators for replacement on the beds. The actuators were sent to their company stores and dealers in early March.

Root Cause

Other

Distribution Pattern

The beds were shipped to the nationwide customers of their one consignee who is located in KS.