CyberKnife System Robotic Radiosurgery System, (Standard Treatment Couch) Synchrony Respiratory Tracking System, a Radiation Therapy Device, Model Numbers: 026429, 028185, Product is manufactured and distributed by Accuray, Inc., Sunnyvale, CA A Radiation therapy device.

FDA Recall Z-0908-2011 · Posted 2011-01-18

Recalling FirmAccuray Inc
Recall StatusTerminated
Product CodeIYE
Date Initiated2009-03-16
Date Terminated2011-01-19

Reason for Recall

The product has the potential for the extension coupling of the Synchrony extension mounting hardware to become loose over time. The assembly weighs approximately 17 pounds; an unexpected contact with users or patients may result in serious injury if the boom descends unexpectedly.

Firm Action

Accuracy Inc., sent an URGENT ADVISORY NOTIFICATION letter, dated March 16, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their system to see if it was affected. Customers were instructed to contact Accuracy Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 or e-mail [email protected] to schedule service.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - USA (nationwide) and the countries of Turkey, Russia, France, Italy, the Netherlands, the UK, Spain, Greece, Switzerland, Germany, India, Taiwan, China, Hong Kong, Korea, Thailand, Malaysia and Japan