Codman Surgical Patties, Part number 801400; SURG PATXRAY 1/2X1/2-200; Lot 6396457

FDA Recall Z-0909-2023 · Posted 2023-01-06

Recall StatusOpen, Classified
Product CodeHBA
Date Initiated2022-12-02
Product Quantity168 US; 3 OUS

Reason for Recall

Integra LifeSciences identified that one lot was released with out-of-specification endotoxin results. The worst-case risk for the use of affected Surgical Patties is potentially a minor transient fever.

Firm Action

On December 2, 2022, the firm notified customers via "URGENT VOLUNTARY MEDICAL DEVICE RECALL" letters. Customers were instructed to immediately remove units of affected product from service. Customers should complete the Acknowledgement Form and return to the recalling company. Once received, Customer Service will contact the customer to aid them in returning the affected product. Should you have any questions regarding these instructions, please contact Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: [email protected]

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, IN, KY, MA, MI, MO, NC, NJ, NY, OH, PA, TX, VA, WV and the countries of Chile, Australia, New Zealand, France, Romania, United Kingdom, and Ireland.