Terminated

Varian Medical Systems, Eclipse Client software versions with build numbers from 7.3.10 through 8.6.15 (i.e. excluding 8.6.17); Model Number: H48; Radiation Treatment Planning System

FDA Recall Z-0911-2011 · Posted 2011-01-18

Recall StatusTerminated
Product CodeMUJ
Date Initiated0010-08-17
Date Terminated2011-07-21
Product Quantity7019

Reason for Recall

Possible incorrect dose calculation: An anomaly in certain Eclipse versions where, under certain specific sequences of user actions, the dose distribution may fail to be re-calculated even though changes have been made which affect dose.

Firm Action

Varian Medical Systems Urgent Medical Device Correction letters were sent to all affected users on August 6, 2010. Letters were also sent to Varian Sales, Marketing and Service organizations to inform them of the recall.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.