Terminated

BD Vacutainer PST Gel and Lithium Heparin (LH); 83 Units; Plus Blood Collection Tubes; Sterile; IVD; REF 367962; 4.5 mL, 13 X 100 mm; BD, Franklin Lakes, NJ USA; Made in USA. Used in hospitals, outpatient clinics and physicians' offices to collect blood from patients for specialized chemistry testing.

FDA Recall Z-0911-2013 · Posted 2013-03-06

Recall StatusTerminated
Product CodeJKA
Date Initiated2012-03-05
Date Terminated2015-07-20
Product Quantity225 million units

Reason for Recall

BD is conducting a recall of the BD Vacutainer Plus Plastic Plasma Separator Tube (PST) with Light Green/BD Homogard Closure due to a deformed top of the tube reservoir reported in two customer complaints.

Firm Action

BD sent advisory letters via UPS dated 3/5/2012 to customers. The letter identified the affected product and the reason for recall. BD has taken steps to modify the existing manufacturing and assembly processes to help reduce the potential for this condition. Questions regarding this issue should be directed to BD Global Technical Services at 1-800-631-0174 or email at www.bd.com/Vascutainer/contact.

Root Cause

Equipment maintenance

Distribution Pattern

Nationwide Distribution