ev3 Primus GPS Biliary Stent System. Balloon Expandable Stent and Delivery System. The stent is made from a 316L stainless steel tube cut into an open lattice design. It is mounted onto a non-compliant balloon catheter. It is deployed and expanded by inflating the balloon.

FDA Recall Z-0912-06 · Posted 2006-05-31

Recalling FirmEv3, Inc
Recall StatusTerminated
Product CodeFGE
Date Initiated2006-03-08
Date Terminated2007-02-03
Product Quantity29

Reason for Recall

A Primus device labeled (pouch and box) as a BXB35-05-17-120 contained a 6mm x 20mm catheter instead of the intended 5mm x 20mm catheter.

Firm Action

ev3 field representatives visited all affected accounts. Product was retrieved a 'Field Action Form' was completed and a copy was left with the account.

Root Cause

Other

Distribution Pattern

Only distributed in US to CA, FL, MA, MD, MS, NY, PA, TN, TX, WA, WV