ACL Futura, ACL Advance and ACL TOP Coagulation Analyzers used with the HemosIL RecombiPlasTin reagent

FDA Recall Z-0915-05 · Posted 2005-06-21

Recall StatusTerminated
Product CodeJPA
Date Initiated2005-05-19
Date Terminated2006-12-06
Product Quantity617 units

Reason for Recall

Patient prothrombin time (PT) may report low for individuals on anticoaglant therapy

Firm Action

Instrumentation Laboratory issued an 'Urgent Product Notification' on 5/19/2005 to Beckman Coulter, Miami FL (DIstributor) to provide the written notification and advise customers to review PT curves an any patient on anticagulant treatment who reports a low result , and manually view the curve where a PT is flagged. A software upgrade will be availabel in July 2005.

Root Cause

Other

Distribution Pattern

Nationwide Canada