AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.

FDA Recall Z-0915-2013 · Posted 2013-03-06

Recalling FirmSteris Corporation
Recall StatusTerminated
Product CodeFLE
Date Initiated2013-02-12
Date Terminated2013-12-05
Product Quantity395 Units

Reason for Recall

The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.

Firm Action

Steris Corporation initiated their recall on February 12, 2013. They will notify customers via onsite visits by STERIS Field Service Technicians to each affected Customer site. For further questions, please call (440) 392-7601.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - USA including Canada and the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, SK, TN, TX, UT, VA, VT, WA, WI, and WV.