Sealapex Express (Part Number: 33639), root canal filling material. Manufactured by SybronEndo, Glendora, CA. Intended as a root canal filling material that is used in conjunction with gutta percha or silver endodontic points.

FDA Recall Z-0917-2010 · Posted 2010-03-02

Recall StatusTerminated
Product CodeKIF
Date Initiated2009-02-23
Date Terminated2012-03-21
Product Quantity1309 units

Reason for Recall

Sybron Dental Specialties is initiating a recall on the Sealapex Xpress due to the presence of crystals in the catalyst. This crystallization may affect proper canal filling by preventing Gutta Percha Points from reaching the bottom of the root canal which may lead to an inadequate seal.

Firm Action

SybronEndo initiated an "Urgent Medical Device Recall" notification dated February 19 , 2009. Consignees were instructed to complete a "return form" and return any affected product in their inventory to the firm. Any product returned will be replaced at no charge. For further information, contact SybronEndo Customer Care at 1-800-346-3636.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution -- United States, Australia, Canada, United Arab Emirates, Belgium, Bulgaria, Germany, Spain, France, UK, Greece, Italy, Kuwait, Lebanon, Portugal, Russia, Sweden, Hong Kong, India, Malaysia and Taiwan.