MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems

FDA Recall Z-0917-2022 · Posted None

Recalling FirmVolcano Corp
Recall StatusOpen, Classified
Product CodeDSA
Date Initiated2022-02-03
Product Quantity1358

Reason for Recall

Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will be replaced with full protective metal brackets.

Firm Action

On 02/03/22, correction notices were sent to customers who were asked to do the following: Notify all affected system users within the facility of this recall. The updated bracket improves protection on all sides of the MM-TSM touch screen that will prevent issues, cosmetic concerns, and maintain device integrity. If you are a distributor or have forwarded affected device(s) to another end user, it is imperative that all end-users with affected devices receive this Important Product Notice. Complete and return the customer response form. Customers with additional questions are encouraged to contact their local recalling firm representative or customer service: Phone: 1-800-228-4728, Option 2, Email: [email protected], Hours of Operation: Monday - Friday 8:00AM 5:00PM PST

Root Cause

Device Design

Distribution Pattern

US: AZ, IL, NV, MI, FL, NC, AK, PA, WI, MN, CA, AR, NJ, TX, OR, NM, SD, CO, KY, LA, DE, MA, MT, MO, VA, AL, NY, WV, SC, IN, GA, IA, OH, UT, NH, MS, TN, OK, ID, MD, HI, RI, WY, DC, WA, KS, NE, ND, CT, VT. OUS: SK, IT, DK, GB, DE, SE, IN, KW, QA, PL, NL, CL, JP, BE, AU, FR, PF, PT, NZ, CA, ES, PK, CY, AT, IL, IE, CH, GR, SA, PS, LB, HK, IS, TH, AR, OM, AE, NO, ID, RO, RS, MM, ZA, SG