Terminated

FUJIFILM Synapse PACS Software: versions 7.0.0 and 7.0.1 - Product Usage: intended for use as a web based application on an off-the shelf PC which meets or exceeds minimum specifications and is networked with a FUJIFILM Synapse PACS server.

FDA Recall Z-0918-2021 · Posted None

Recall StatusTerminated
Product CodeLLZ
Date Initiated2020-12-11
Date Terminated2021-11-17
Product Quantity1 unit

Reason for Recall

The software does not update measurements and calculations in the Clinical Reporting Application (CRA) when the ventricular trace is changed in the study by a different user.

Firm Action

On 12/11/2020 Fujifilm sent consignee notifications via FedEx Priority Overnight stating that there are issues with Synapse PACS Software Version 7.0.000 and 7.0.100. Fujifilm provided a work around until a hotfix is scheduled and completed.

Root Cause

Software change control

Distribution Pattern

US Nationwide distribution in the state of MD.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.