Roche VANC2, Online TDM Vancomycin, for use with the Roche/Hitachi Cobas c 501analyzer; Catalog No.04491050190.

FDA Recall Z-0919-2007 · Posted 2007-06-09

Recall StatusTerminated
Product CodeLEH
Date Initiated2007-04-30
Date Terminated2007-12-17
Product Quantity662

Reason for Recall

Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.

Firm Action

Consignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.

Root Cause

Other

Distribution Pattern

Nationwide.