Terminated

MUSE Cardiology Information System, Model Numbers: a) 3101011-040, b) 3101010-401, c) 3101010-009, d) 3101010-008

FDA Recall Z-0919-2022 · Posted None

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeDQK
Date Initiated2022-03-16
Date Terminated2026-05-04
Product Quantity147 units

Reason for Recall

Two scenarios may cause edits to measurements and diagnosis statements can be lost after a test is Signed in the MUSE NX web client.

Firm Action

GE Healthcare notified its consignees by letter titled URGENT MEDICAL DEVICE CORRECTION on March 16, 2022. The letter explained the two scenarios which would produce the loss of edits and provided recommendations and a workaround. The firm will correct all affected products at no cost to you. The firm will be issuing a software patch. A GE Healthcare representative will contact affected customers to arrange for the correction. If you have any questions or concerns regarding this notification, contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative

Root Cause

Software Manufacturing/Software Deployment

Distribution Pattern

US and Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, England, France, Germany, Ireland, Italy, Korea, Malaysia, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey