Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190.
FDA Recall Z-0920-2007 · Posted 2007-06-09
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeLEG
Date Initiated2007-04-30
Date Terminated2007-12-17
Product Quantity992
Reason for Recall
Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.
Firm Action
Consignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.
Root Cause
Other
Distribution Pattern
Nationwide.