Fresenius Bicarbonate Jug Adapter to the 2008 Series Hemodialysis machine Part Number: 650103. The adapter is a socket that accepts the 2008 machine spike to draw concentrate from the container.
FDA Recall Z-0920-2014 · Posted 2014-02-03
Recalling FirmFresenius Medical Care Holdings, Inc.
Recall StatusTerminated
Product CodeKDI
Date Initiated2013-12-17
Date Terminated2021-05-18
Product Quantity703 units
Reason for Recall
Bicarbonate Jug Adapter distributed prior to
510(k) approval
Firm Action
Fresenius notified customers by telephone and Formal notification on 12/17/13 via the Urgent Bicarbonate Jug Adapter Recall letter sent certified mail with signature confirmation and fax-back form.
Customers were instructed to examine their stock immediately to determine whether they have any of the affected parts on hand. If customers have the affected parts, they are instructed to discontinue use and place all Bicarbonate Jug Adapters in a secure area for return to Fresenius Medical Care Renal Therapies Group, LLC (FMC-RTG). Customers instructed to contact their Fresenius Medical Care Technical Services Team for instructions on how to return the recalled product.
Root Cause
No Marketing Application
Distribution Pattern
Nationwide Distribution.