IMS Patient Scale used on Sunrise Medical's Hoyer Presence and Stature Patient Lifts (Cradle), Models: HOY-4pt-WSC and HOY-StatureWSC

FDA Recall Z-0921-2007 · Posted 2007-06-19

Recall StatusTerminated
Product CodeFRW
Date Initiated2007-01-30
Date Terminated2008-03-02
Product Quantity149 units

Reason for Recall

Patient Scale may become detached, which can cause a patient to fall and result in injury.

Firm Action

The scale manufacturer, IMS, in conjunction with the lift manufacturer, Joerns Healthcare (previously Sunrise Medical), will initiate a field fix strategy consisting of shipping replacement 4-point cradles without scales, for use until a replacement can be designed, tested and validated. Joerns will utilize their Customer Service Help Line (1-800-826-0270) on behalf of IMS, to inform customers of the voluntary field correction, address any questions regarding the IMS scale field fix strategy and to provide additional information on return of affected IMS Patient Scales. Customers were notified by telephone beginning on January 30, 2007.

Root Cause

Other

Distribution Pattern

Worldwide