SOMATOM Perspective, Computed tomography x-ray system
FDA Recall Z-0921-2017 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2016-11-22
Date Terminated2017-05-30
Product Quantity13 systems
Reason for Recall
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
Firm Action
Siemens mailed on December 01, 2016 a Customer Safety Notice to affected customers to inform them of a corrective action that is being performed to prevent a possible hazard to persons and equipment. The notice explained the underlying issue, the effect the issue has on the system, the associated risks and actions they will take to correct the issue.
In addition, customers will receive a Customer Safety Notice via Hand Delivery from the Siemens Customer Service Engineer at the scheduled appointment for the corrective installation.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Distributed to: IL IL GA SC SD OK NY MI IN MD AR