Terminated

Cobas infinity central lab / cobas infinity core license-a Calculator/Data Processing Module, For Clinical Use Material Number: 07154003001

FDA Recall Z-0922-2020 · Posted None

Recall StatusTerminated
Product CodeJQP
Date Initiated2019-10-08
Date Terminated2020-10-15
Product Quantity20

Reason for Recall

Cobasinfinity laboratory solution Version 2.4.1 through Version 2.5.4 Using a CommServer Driver - Incorrect Alarm Mapping

Firm Action

Roche issued Urgent Medical Device Correction TP-00791 mailed via UPS Ground (receipt required) on 10/8/2019. The letter identifies the problem, health risk and action to take: Complete the attached fax form (TP-00792) and fax or email it according to the instructions on the form. " File this Urgent Medical Device Correction (UMDC) for future reference Roche will resolve this issue in a future software update. In the next couple of weeks, a Roche Representative will be contacting you to adjust alarm code mapping as needed per your specific configuration. Contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-526-2272 if you have questions. The UMDC will also post to the diagnostics.roche.com website

Root Cause

Software design

Distribution Pattern

US Nationwide in the states of AL, CA, GA, IA, IL, KY, MI, MO, MT, NJ, NY, OR, TN