ARTIS One Angiographic X-Ray System

FDA Recall Z-0922-2025 · Posted 2025-01-29

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2025-01-02
Product Quantity1140

Reason for Recall

A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

23 Distributed in the US