Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
FDA Recall Z-0923-2026 · Posted 2025-12-22
Advertisement
Recalling FirmAGFA Healthcare Corp.
Recall StatusOpen, Classified
Product CodeIZF
Date Initiated2025-11-18
Product Quantity35
Reason for Recall
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro
exams. This can happen in case the framerate is increased on the NX workstation when ABS is not
enabled.
Firm Action
AGFA NV notified Customers via a letter titled "IMPORTANT PRODUCT INFORMATION LETTER / DR 800", included Product Name, Symptom, Cause, Actions to be taken customer and manufacturers respectively: "Always enable ABS for pulsed fluoro exams.
Actions being taken by the manufacturer:
Agfa/local business partner will contact you to install a new generator software."
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
U.S. Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.