Medtronic Gastro pH in a Box System, Model 43A008, Catalog Number 9043A0081

FDA Recall Z-0924-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeFFX
Date Initiated2004-04-28
Date Terminated2005-11-21
Product Quantity42 units

Reason for Recall

Inadequate wire insulation inside the plugs of the power cords of the devices may cause the power cords to short.

Firm Action

Letters dated April 28, 2004 were sent to the consignees. Enclosed with the letter to each consignee were replacement power cord(s). The letters instruct consignees to replace the current power cord by following instructions enclosed with the letter.

Root Cause

Other

Distribution Pattern

The devices were distributed nationwide in the United States.