ABX PENTRA Urine Control L/H. The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the labeling, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA chemistry analyzers. Each control level is provided in one vial of 10 ml. The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.
FDA Recall Z-0924-2014 · Posted 2014-02-03
Recalling FirmHoriba Instruments, Inc dba Horiba Medical
Recall StatusTerminated
Product CodeJJY
Date Initiated2013-12-18
Date Terminated2014-10-09
Product Quantity531
Reason for Recall
HORIBA Medical is informing all ABX PENTRA Urine Control L/H customers running the Creatinine assay
on ABX PENTRA Chemistry Analyzers that the Target Range for Creatinine has been updated. The ABX PENTRA Urine Control L/H lot specific annexes have been updated to
include revised acceptable Target Ranges for Creatinine.No other analyte values
are affected.
Firm Action
HORIBA Medical sent an Urgent Safety Notice dated December 18, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to update their Quality Control procedures to include the new Creatinine Target
Ranges for the ABX PENTRA Urine Control L/H Reference #: A11A01674 and Lot #s: 1207001, 1303601, and 1300701. Customers were instructed to complete and return the enclosed response form to verify their understanding of the Field Safety Notice. Customers with questions were instructed to contact their local HORIBA Medical representative.
For questions regarding this recall call 949-453-0500.
Root Cause
Pending
Distribution Pattern
Worldwide Distribution - USA (nationwide) and Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K110530ABX PENTRA CREATININE 120 CP; ABX PENTRA MULTICAL; ABX PENTRA N CONTROL; ABX PENTRA P CONTROL; ABX PENTRA URINE CONTROLHORIBA ABX SAS · 2012-04-30K110137ABX PENTRA ENZYMATIC CREATININE CP, ABX PENTRA MULTICAL, ABX PENTRA N CONTROL, ABX PENTRA P CONTROL, AND ABX PENTRA URINHORIBA ABX SAS · 2011-08-10
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.