Neodent Tapered Guided Surgery Drill, Article No. 103.436. UDI 07899878030041. The device is a dental burr.
FDA Recall Z-0924-2020 · Posted 2020-01-30
Advertisement
Recalling FirmJjgc Industria E Comercio De Materials Dentarios Sa
Recall StatusTerminated
Product CodeEJL
Date Initiated2019-12-23
Date Terminated2021-06-01
Product Quantity2
Reason for Recall
The laser engraving of the product incorrectly identifies the 4.3 Tapered Guided Surgery Drill as the "4.3+ Tapered Guided Surgery Drill".
Firm Action
Customers were notified of the recall on 12/23/2019 via telephone and email.
Customers were instructed to do the following:
- If the clinician has already used the drill for osteotomy and successfully concluded the installation of
the implant, additional patient follow-up not required. The implant will behave as normally
expected.
- Identify and segregate the affected units, opened or unopened, in your stock.
- If the product is still in your inventory, return it to the attention of "Action 0002/2019 Product Removal" for credit as indicated on ANNEX 1: Customer Confirmation Form.
- If the product cannot be located, indicate this on the Customer Confirmation Form.
- For either case, complete and return the Customer Confirmation Form to "Action 0002/2019 Product Removal"
Root Cause
Under Investigation by firm
Distribution Pattern
Distributed to consignees in CA and PA.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.