SmartSite Needleless Connector, Model No. 2000E The SmartSite Needleless Connectors are sterile, individually packaged connectors. The connectors are used to administer fluid and medications. The needle free connector allows the user to add medication into the primary line without the use of a needle.

FDA Recall Z-0925-2014 · Posted 2014-02-03

Recalling FirmCareFusion 303, Inc.
Recall StatusTerminated
Product CodeFPA
Date Initiated2014-01-22
Date Terminated2015-08-06
Product Quantity3,500,000 units (2,136,200 units in US; 1,404,700 units foreign)

Reason for Recall

CareFusion is recalling the SmartSite Needlefree Connector, model number 2000E, because of connection issues. The affected SmartSite connector lots may unintentionally disconnect from a female luer, may be difficult to disconnect from a female luer, or may fail to disconnect from a female luer once attached. Leakage may also be observed if the connector disconnects from the female luer during in

Firm Action

The firm, CareFusion, sent an "URGENT: Medical Device Recall Notification" letter dated January 22, 2014 to customers. The letter described the product, problem and actions to be taken. Customers were instructed to discontinue use of the listed lot numbers, contact CareFusion for a replacement of the affected lots, and if you have inventory of the listed products immediately complete and return to CareFusion the enclosed, pre-addressed and postage paid, Recall Response Card. All recalled product should be returned directly to the distributor from whom it was purchased. If you have any question, contact your distributor directly or call the CareFusion Support Center at numbers below. For recall related questions, customers are instructed to contact CareFusion Support center at (888) 562-6018. For adverse event reports, customers are instructed to contact Customer advocacy at (888) 812-3266 or email [email protected]. For technical questions regarding the Alaris System, customers are instructed to contact technical support at (888) 812-3229.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution: US (nationwide) including Puerto Rico; and internationally to: Australia, Canada and New Zealand.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.