NeatNick Sweeping Action Heel Lancet -Full-Term- Item Number: 1030085 (box of 100) and Item Number: 1030086, case of 1,000 (10 boxes of 100)

FDA Recall Z-0926-2007 · Posted 2007-06-09

Recalling FirmHawaii Medical LLC
Recall StatusTerminated
Product CodeFMK
Date Initiated2007-05-11
Date Terminated2008-07-22
Product Quantity286,376 units

Reason for Recall

Lancet may misfire and an inadequate blood sample may result; additionally, the blade may not retract after use and result in a needle stick to the user/patient

Firm Action

Hawaii Medical notified the initial distributor and end users by telephone and follow-up Recall Notification Letter on May 15, 2007.

Root Cause

Other

Distribution Pattern

Nationwide