DPOC-20, Direct Peroral Cholangioscopy Balloon, G54951, 20 mm/17 mL, Rx Only, Cook Medical, Winston-Salem, NC 27105 This device is intended to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. Also, used for endoscopic removal of biliary stones.

FDA Recall Z-0926-2011 · Posted 2011-01-19

Recalling FirmCook Endoscopy
Recall StatusTerminated
Product CodeFGE
Date Initiated2010-12-23
Date Terminated2011-08-04
Product Quantitysee line 1.

Reason for Recall

There is a risk of gastrointestinal perforation or air embolism due to balloon deflation difficulty, resulting in poor patient outcomes.

Firm Action

The firm, Cook Medical, sent an "URGENT: PRODUCT RECALL" letter dated January 7, 2011, to all customers. The letter described the product, problem and the actions to be taken by the customers. The customers were instructed to quarantine any affected product and return to Cook Medical. If they are distributors, they were instructed to contact their customers, direct them to quarantine any affected product and coordinate the return of the affected product to Cook Medical. The customers were also instructed to complete and return the bottom of the letter and send to Cook Medical as well as the PRODUCT RECALL EFFECTIVENESS SURVEY via fax at 812-339-7316 or scanned and emailed to [email protected]. Note: All non-responding customers will receive a total of three (3) recall notifications ( two (2) follow up requests for a response in addition to the initial recall notification). Should you have any questions, please contact the Customer Relations Department at Cook Medical for more information. Please use [email protected] or call toll free at 1-800-457-4500, press 4 then enter extension 2146.

Root Cause

Other

Distribution Pattern

Worldwide distribution: USA including states of: AL, CA, FL, IL, IN, LA, MA, NC, NY, OH, PA, SC, TN, TX, WA, and WI; and countries of: Belgium, Canada, Germany, Hong Kong, and the Netherlands..