NeatNick Sweeping Action Heel Lancet - Preemie- Item Number: 1030083 (box of 100) and Item Number: 1030084, case of 1,000 (10 boxes of 100)
FDA Recall Z-0927-2007 · Posted 2007-06-09
Advertisement
Recalling FirmHawaii Medical LLC
Recall StatusTerminated
Product CodeFMK
Date Initiated2007-05-11
Date Terminated2008-07-22
Product Quantity58,058 units
Reason for Recall
Lancet may misfire and an inadequate blood sample may result; additionally, the blade may not retract after use and result in a needle stick to the user/patient
Firm Action
Hawaii Medical notified the initial distributor and end users by telephone and follow-up Recall Notification Letter on May 15, 2007.
Root Cause
Other
Distribution Pattern
Nationwide