Anestar and Anestar Plus Anesthesia System.

FDA Recall Z-0929-06 · Posted 2006-06-02

Recalling FirmDatascope Corp
Recall StatusTerminated
Product CodeBSZ
Date Initiated2006-03-17
Date Terminated2006-08-01
Product Quantity337 UNITS

Reason for Recall

The Anestar and Anestar Plus Anesthesia System may produce periods of high Positive End-Expiratory Pressure (PEEP), in both manual and mechanical ventilation mode due to an issue with the Applied Pressure Limit (APL) valve.

Firm Action

Field correction communications were sent 3/17/2006 to all accounts via certified mail, return receipt requested, signature required.

Root Cause

Other

Distribution Pattern

The units have been distributed to distributors and end-users such as hospitals and clinics nationwide. There are 3 foreign accounts in Australia, El Salvador, and Venezuela.