Somatom Definition AS, Model number 8098027. Computed tomography x-ray system

FDA Recall Z-0929-2009 · Posted 2009-01-23

Recall StatusTerminated
Product CodeJAK
Date Initiated2008-12-09
Date Terminated2009-09-23
Product Quantity42 units

Reason for Recall

possible improper installation of tilt transport lock/tilt limiter

Firm Action

The recalling firm received a letter dated 11/25/08 from the manufacturer informing them of the problem. As a result, the recalling firm has issued Update Instruction CT065/08/S to have the transport lock/tilt limiter on the device visually inspected. The tilted scan modes have been disabled until the modified lock/tilt limiter has been installed.

Root Cause

Labeling Change Control

Distribution Pattern

Nationwide Distribution --- including states of AZ, CA, CO, FL, IL, KS, KY, MA, MI, MN, MO, NE, NY, OH, PA, SC, VA, and WY.