Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Atellica CH Enzymatic Hemoglobin A1c (A1c_E/A1c_H), Siemens Material Number 11097536

FDA Recall Z-0929-2022 · Posted None

Recall StatusOpen, Classified
Product CodePDJ
Date Initiated2022-02-09
Product Quantity6606

Reason for Recall

Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.

Firm Action

Urgent Medical Device Correction notification letters dated 2/9/22 were sent to customers. Actions to be Taken by the Customer Please review this letter with your Medical Director. Perform the instructions provided in the Additional Information section below. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received these products. Additional Information If your laboratory has multiple Atellica CH 930 Analyzers, separate the assays as follows: -LITH_2 from CA_2 and TBil_2 -A1c_E/A1c_H from CA_2 - CA_2 from UN_c (urine only); UN_c serum and plasma are not impacted and can be run on the same analyzer as CA_2 -Fruc from AAT, IgA_2, IgM_2 and Theo If you choose not to separate the assays as indicated above, batch testing of A1c_E, Fruc, LITH_2, and UN_c (urine only) may be considered. Note: An RPC2 wash mitigation must be initiated after completion of A1c_E, Fruc, LITH_2, and prior to UN_c (urine only). Any of the following will initiate the RPC2 wash: -After the Atellica CH 930 Analyzer has been in standby for 12 minutes. - After completion of any Open Channel assay. - Restarting the Atellica CH 930 Analyzer. Refer to the Atellica Solution Online Help for instructions on system restart. If you have any questions, contact your local Siemens Healthineers technical support representative.

Root Cause

Device Design

Distribution Pattern

Global Distribution. US Nationwide.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑