Harmony vLED Surgical Lighting System, one unit packaged per box The Harmony vLED Surgical Lighting System is a variable pattern, variable intensity surgical lighting fixture designed to provide visible illumination of the surgical field or the patient for the operating room staff.

FDA Recall Z-0930-2015 · Posted 2015-01-07

Recalling FirmSteris Corporation
Recall StatusTerminated
Product CodeFST
Date Initiated2014-11-26
Date Terminated2015-07-11
Product Quantity140 units

Reason for Recall

STERIS has learned that groups of LED lights (modules) contained within the vLED light head may illuminate inconsistently.

Firm Action

STERIS sent an Urgent Voluntary Field Correctionn Notice to all affected consignees via FedEx. The Consignee Notification mailing was initiated the week of December 15, 2014. Customers were instructed that their local STERIS representative would contact their facility to arrange for the installation of updated software. Customers with questions were instructed to call 1-800-548-4873. For questions regarding this recall call 440-392-7231.

Root Cause

Software Design Change

Distribution Pattern

Worldwide Distribution - USA including Alabama, Alaska, Arkansas, California, District of Columbia, Florida, Illinois, Indiana, Kansas, Louisiana, Mississippi, Nevada, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Utah, and Virginia and Internationally to Canada, China, Greece, Indonesia, Republic of Korea, Romania, Russian Federation, Taiwan, and Viet Nam.

Related FDA 510(k) Records