Baxter Colleague Single Channel Volumetric Infusion Pump. Baxter Healthcare Corporation, Medication Delivery Division. Product Code: 2M8161
FDA Recall Z-0931-2011 · Posted 2011-01-21
Advertisement
Recalling FirmBaxter Healthcare Corp.
Recall StatusTerminated
Product CodeFRN
Date Initiated2010-12-01
Date Terminated2011-05-13
Product Quantity1 pump
Reason for Recall
While the infusion pump was being serviced at the Baxter Service Center, it was inadvertently routed to the packaging area before the final inspection.
Firm Action
Baxter telephoned the one customer on December 1, 2010, informing them that that an outgoing Quality Control Evaluation was not performed on their Colleague pump serial number 15030799CC during the last service. Since this was not conducted on this pump, the device may not function as designed and could lead to an interruption of therapy. Baxter requested that they locate the pump and remove it from use as soon as possible. Baxter will then send a shipping container for retrieval, reinspection and return of the device back to the customer at no cost.
For questions regarding this recall call (800) 437-5176.
Root Cause
Employee error
Distribution Pattern
Texas
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.