Terminated

Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process.

FDA Recall Z-0931-2013 · Posted 2013-03-10

Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeIYE
Date Initiated2012-12-19
Date Terminated2021-07-14
Product Quantity164

Reason for Recall

The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.

Firm Action

Elekta sent an "IMPORTANT FIELD SAFETY NOTICE" dated December 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The IFSN informs the user in the TPS, to give the different volumes different structure names. After importing the patient data set from the TPS, make sure that the structure contours are correct. Users are requested to return the signed acknowledgement. Contact your local Elekta office for questions regarding this notice.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution-DC, Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, FL, GA, IL, IA, KY, MD, MA, MI, MN, MS, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WA, WV, and WI.