Terminated

Yukon Polyaxial Screw Size 4.0x26mm (catalog number 7601-04026), component of YUKON Spinal System

FDA Recall Z-0931-2021 · Posted 2021-01-27

Recalling FirmK2M, Inc
Recall StatusTerminated
Product CodeNKG
Date Initiated2020-12-14
Date Terminated2023-07-11
Product Quantity248 units/1 each

Reason for Recall

The Yukon polyaxial screw used in the Yukon Spinal System was incorrectly laser marked with size 5.0 mm (diameter) & 24 mm (length).

Firm Action

On 12/14/2020, Stryker Spine US Field Operations were notified of the action by teleconference and e-mail and instructed to identify and return all Lot KRBX screws to Stryker Spine. Affected Stryker Branches/Agencies, and hospitals that had previously been consigned product, will be issued a formal recall notification letter and product return documentation.

Root Cause

Employee error

Distribution Pattern

Domestic: Nationwide Foreign: Australia, Canada, Netherlands, Singapore