Terminated

Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040.

FDA Recall Z-0932-2007 · Posted 2007-06-30

Recall StatusTerminated
Product CodeDTQ
Date Initiated2005-08-01
Date Terminated2008-02-23
Product Quantity1982 total for both models distributed

Reason for Recall

The tube clamp mechanism on the roller pump may not open when the rotation knob is turned, making it difficult to remove and/or insert tubing.

Firm Action

The firm is visiting the consignees as complaints are received and making corrections. Corrections were made in August 2005 and April 2006 to date. Other units will be repaired if complaints are received and the firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and appropriate action to take should it occur.

Root Cause

Other

Distribution Pattern

Worldwide; USA, Australia, Belgium, Canada, Dominician Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Korea, Malaysia, Mexico, Pakistan, Republic of Georgia, Singapore, South Africa, Taiwan, Thailand and United Arab Emirates.