Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Solanas Posterior Stabilization System Part Number: 63920. Solanas Posterior Stabilization System facilitates the surgical correction of spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. The Solanas Posterior Stabilization System is a spinal fixation system intended to improve stability of the cervical and thoracolumbar area of the spine (C1-T3).

FDA Recall Z-0932-2012 · Posted 2012-01-31

Recalling FirmAlphatec Spine, Inc.
Recall StatusTerminated
Product CodeKWP
Date Initiated2010-11-01
Date Terminated2012-01-31
Product Quantity27 units

Reason for Recall

These lots were not manufactured correctly in that the instrument does not have the durability required to repeatedly perform over its full service life and, therefore, may experience breakage in the tips.

Firm Action

Alphatec Spine contacted all affected customers via telephone. The communication included discussion of the product, problem, and actions to be taken by the customers. Customers were instructed to remove the affected product from inventory and to return the product to the firm. Arrangements were made for replacement product to be sent to the customers. Contact the firm at 1-800-922-1356 for questions regarding this recall.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution-including the states of CA, CT, FL, GA, NC, NE, NJ, TX, and UT.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑