Hydroxybutyrate LiquiColor Tests - includes 3 bottles (1 of each size, 3mL, 8.5mL, and 50mL)
FDA Recall Z-0932-2022 · Posted None
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusTerminated
Product CodeJIN
Date Initiated2022-03-08
Date Terminated2023-03-20
Product Quantity45 ea
Reason for Recall
Product was improperly store which may cause a delay in results.
Firm Action
Consignees were notified via first class mail on March 08, 2022. Consignees are responsible for initiating sub-recalls, where necessary.
Medline Industries, LP. has provided each consignee with a letter and a Destruction form to record their response to the recall notification. Each Response received will be recorded and Medline Industries, LP. will send subsequent letters to consignees that have not provided a response to the recall notification in the allotted time.
Each consignee was provided with a destruction form to complete and return. The consignees are responsible to destroy their product per their local and state regulations. No product will be returned to Medline.
Root Cause
Storage
Distribution Pattern
US distribution only - AL, CA, FL, GA, IL, IN, LA, MO, MS, NC, NE, NY, OH, PA, VA, and WA
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.