Volar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W, VBEAR-3-7N, VBEAR-5-7-N, VBEAR-3-7S, VBEAR-5-7S, VBEAR-7-7S, VBEAR-3-7W, VBEAR-5-7W. Intended to be used as an aid to the treatment of certain types of fractures, non-unions or osteotomies.

FDA Recall Z-0932-2024 · Posted 2024-01-31

Recalling FirmTriMed Inc.
Recall StatusOpen, Classified
Product CodeHRS
Date Initiated2023-11-28
Product Quantity270

Reason for Recall

Plates, part of a wrist fixation system, are affixed during surgery with bearings that may pop out from distal peg holes.

Firm Action

On 11/28/23, recall notices were emailed to customers who were asked to do the following: 1) Stop further use of the device and return any unused products to the recalling firm. 2) Provide a copy of this recall notice to the hospital/surgeon/physician that is using the device. 3) Complete and return the response form to [email protected] Contact [email protected] for any additional questions.

Root Cause

Labeling design

Distribution Pattern

US: FL, MI, AL, NY, CA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.