Synchron Systems Drug Calibrator 1 (Carbamaxepine, Phenobarbitol, Phentoin, Theophylline, and Valproic Acid).
FDA Recall Z-0935-04 · Posted 2004-07-20
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeKLT
Date Initiated2003-06-11
Date Terminated2004-05-10
Product Quantityunknown
Reason for Recall
Product was found to produce higher recoveries by about 7-12% compared to other lots.
Firm Action
Users were requested to discard and ask for replacement of the lot in question by letter.
Root Cause
Other
Distribution Pattern
Nationwide and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.