Synchron Systems Drug Calibrator 1 (Carbamaxepine, Phenobarbitol, Phentoin, Theophylline, and Valproic Acid).

FDA Recall Z-0935-04 · Posted 2004-07-20

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeKLT
Date Initiated2003-06-11
Date Terminated2004-05-10
Product Quantityunknown

Reason for Recall

Product was found to produce higher recoveries by about 7-12% compared to other lots.

Firm Action

Users were requested to discard and ask for replacement of the lot in question by letter.

Root Cause

Other

Distribution Pattern

Nationwide and Canada