Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

LOCATOR Implant Attachment System. The LOCATOR Implant Attachment System is a universal hinge, resilient attachment for endosseous implants. The LOCATOR Implant Attachment System is designed for use with overdentures or partial dentures, retained in whole or in part, by endosseous implants in the mandible or maxilla. FDA 510(k): K994257, K072878

FDA Recall Z-0935-2014 · Posted 2014-02-06

Recalling FirmZest Anchors LLC
Recall StatusTerminated
Product CodeDZE
Date Initiated2014-01-16
Date Terminated2014-05-15
Product Quantity3,130

Reason for Recall

Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend

Firm Action

Zest Anchors, LLC initiated this recall by sending customers recall notification letter on 01/16/2014 via FedEx overnight. The customer notification letter, titled, "IMPORTANT: MEDICAL DEVICE VOLUNTARY RECALL NOTICE", informed customers of the recall, reason for voluntary recall, product information with part numbers and lot numbers, actions to be taken, and contact information. The customer notification letter was accompanied by a response form titled "MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement & Receipt Form".

Root Cause

Unknown/Undetermined by firm

Distribution Pattern

Distributed in Ohio and Switzerland.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑