Abbott Latex-Free Primary I.V. Plumset Convertible Pin, 95 inch with Injection Site and Secure Lock, No. 19020; Made in Costa Rica, Abbott Laboratories, North Chicago, IL 60064
FDA Recall Z-0936-03 · Posted 2003-06-17
Advertisement
Recalling FirmAbbott Laboratories HPD/ADD
Recall StatusTerminated
Product CodeFPA
Date Initiated2003-05-13
Date Terminated2003-11-20
Product Quantity4,656 units
Reason for Recall
The I.V. administration set incorrectly lists the priming volume from the cassette inlet to distal male adapter as 13.5 mL instead of 11.3 mL
Firm Action
Recall letter dated 5/12/03 sent to the account on 5/13/03, informing them of the labeling discrepancy and requesting the return of any inventory of the affected lot to Abbott for credit.
Root Cause
Other
Distribution Pattern
Miami, FL