Terminated

Cervical Extension Machine for Physical Therapy

FDA Recall Z-0936-2015 · Posted 2015-01-07

Recalling FirmMedX Holdings, Inc.
Recall StatusTerminated
Product CodeIKK
Date Initiated2013-12-03
Date Terminated2016-09-29
Product Quantity4

Reason for Recall

Review of DHRs revealed that HiPot, Continuity, load and friction re-testing was not completed on these devices during manufacturing.

Firm Action

Medx began contacting its consignees December 30, 2013, to schedule testing on the devices at their facility. The technician has been scheduled to complete required testing of these devices in the field. For questions regarding this recall call 322-351-2005.

Root Cause

Unknown/Undetermined by firm

Distribution Pattern

Worldwide Distribution - USA including MO, IL, ID, and MN and Internationally to Germany and South Korea.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →