Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Argyle Suction Tubing (Product # 8888301515) Argyle Suction Tubing is an extruded tube used as an extension for different types of devices in an operating room, bonded with 2 female sure grip connectors at each end of the tube to facilitate the plug/ insertion of the device and also includes a male connector that can be used for devices designed with female connectors, This product is sterile, non-conductive and latex free.

FDA Recall Z-0936-2018 · Posted 2018-03-05

Recall StatusTerminated
Product CodeGAZ
Date Initiated2017-10-24
Date Terminated2020-10-26
Product Quantity58,954

Reason for Recall

Incorrect packaging. In one lot of Argyle Suction Tubing, the pouch may be incorrectly labeled as the Salem Sump Dual Lumen Stomach Tube. This product may have been distributed within the case of Argyle Suction Tubing.

Firm Action

An URGENT MEDICAL DEVICE RECALL letter was issued to customers on October 24, 2017. This letter notified customers of the products subject to the recall and stated the following: The primary package labeling (Label closest to the product) may be incorrect. Customers can identify the affected product by consulting Attachment B. Please quarantine and return any unused products of the item codes and lot detailed above. Unused products from the affected item code and lots should be returned as described in the Required Actions section below. If you have distributed the Argyle" Suction tubing listed above, please promptly forward the information from this letter to those recipients. All unused products from the affected item codes and lots must be returned. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.

Root Cause

Packaging process control

Distribution Pattern

USA (nationwide) Distribution

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑