Terminated

Sensis Vibe System, Model Number 11007642, with software version VD10B.

FDA Recall Z-0936-2019 · Posted None

Recall StatusTerminated
Product CodeDQK
Date Initiated2018-11-29
Date Terminated2020-05-07
Product Quantity1

Reason for Recall

A software error may result in a system crash. The system must be restarted before the clinical procedure can be continued. The ablation treatment must be performed with a different system or without the use of the interface.

Firm Action

Urgent Medical Device Correction letters were sent to customers on 11/29/18.

Root Cause

Software design

Distribution Pattern

The products were distributed to the following US states: NJ.